Beaver Visitec recalls

6 recalls on record · latest: March 12, 2019

Official U.S. recall history for Beaver Visitec, compiled from official government records.

Get recall alerts

Free email alert whenever Beaver Visitec has a new recall — straight from official government data. Unsubscribe anytime.

Beaver Xstar Slit Knife 2 recalled over labeling errors

March 12, 2019 · Medical device recalls Moderate risk Mislabeled. The package labeling indicates a size 2.4mm, 45 degree Double Bevel…

Beaver Visitec recalls Micro Knife 5.0mm 30¿, Part Number 0001522 The Beaver¿ Micro Knife 5.0mm 30¿ is a sharp pointed tip intended for ophtha…

February 19, 2019 · Medical device recalls The product contains a misprinted expiration date on the peel pack. Both were p…

Beaver Visitec recalls BVI Visitec Soft Tip Cannula, 0.8MM .50 x 33mm (25G x 1.3 in). The BVI Soft Tip Cannula is intended to irrigate and/or…

November 30, 2018 · Medical device recalls Moderate risk Certain lots are missing a protective sheath component over the silicone tip, w…

Beaver Visitec recalls Beaver Accu-Temp High Temperature Cautery, 2 inch Shaft with Fine Tip, Model Number 8445000 Product Usage: The indicati…

July 19, 2018 · Medical device recalls Moderate risk A small number of devices in this lot may have a melted cap and a hole formed t…

Slit Knife 2 recalled over labeling errors

June 18, 2018 · Medical device recalls Moderate risk Mislabel of product 373726 Lot 6001700 with 374891 Lot 6001701 on individual pa…

Beaver Visitec recalls Safety Sideport Knife 1.0mm 45¿, REF 378210 Product Usage: Indicated to make stab incisions during ophthalmic surgery

May 14, 2018 · Medical device recalls Moderate risk Blades are loose in handles and may fall out due to non cured epoxy