Medical device recalls Moderate risk

3M Company - Health Care Business product recalled over fire hazard

Recall date
January 14, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
3M Company - Health Care Business recalls 3M(TM) Surgical Clipper by Remington(R) REF 9602 The 3M Surgical Clipper Classic, 9602, is intended to remove body hair…
Recall number
Z-0791-2019
FDA classification
Class II
Brand / firm
3M Company - Health Care Business
Sold / distributed
US Distribution and countries Brazil and Canada.

Why it was recalled

3M has recently received reports involving the 3MTM Surgical Clipper by Remington¿, Model 9602, overheating in mobile/field EMS situations while charging from a modified sine wave power inverter.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

3M(TM) Surgical Clipper by Remington(R) REF 9602 The 3M Surgical Clipper Classic, 9602, is intended to remove body hair and head hair from patients in preparation for any medical procedure requiring hair removal.