3M Company - Health Care Business product recalled over fire hazard
- Recall date
- January 14, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- 3M Company - Health Care Business recalls 3M(TM) Surgical Clipper by Remington(R) REF 9602 The 3M Surgical Clipper Classic, 9602, is intended to remove body hair…
- Recall number
- Z-0791-2019
- FDA classification
- Class II
- Brand / firm
- 3M Company - Health Care Business
- Sold / distributed
- US Distribution and countries Brazil and Canada.
Why it was recalled
3M has recently received reports involving the 3MTM Surgical Clipper by Remington¿, Model 9602, overheating in mobile/field EMS situations while charging from a modified sine wave power inverter.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
3M(TM) Surgical Clipper by Remington(R) REF 9602 The 3M Surgical Clipper Classic, 9602, is intended to remove body hair and head hair from patients in preparation for any medical procedure requiring hair removal.
Get recall alerts
Free email alert whenever 3M Company - Health Care Business has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: 3M Company - Health Care Business