3M Company - Health Care Business recalls Block Versions Used by Dental Offices often referred to as chairside milling: Lava Ultimate Restorative for CEREC, Com…
- Recall date
- June 15, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2052-2015
- FDA classification
- Class II
- Brand / firm
- 3M Company - Health Care Business
- Sold / distributed
- Worldwide Distribution - US: Nationwide and the countries of: Australia, India, Japan, Korea, New Zealand, Singapore, Thailand, Vietnam, Canada, Bulgaria, Czech Republic, Estonia, Hungary, Israel, Latvia, Lithuania, Poland, Romania, Slovakia, Switzerland, Turkey, China Hong Kong, Taiwan, Brazil, Co…
Why it was recalled
3M is initiating a Field Correction to remove the crown indication from all Lava Ultimate products. These products will continue to be indicated for onlays, inlays and veneers.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Block Versions Used by Dental Offices often referred to as chairside milling: Lava Ultimate Restorative for CEREC, Compatible with Sirona mill 2912x, 2914x, 3312x, 3314x. Lava Ultimate Restorative for E4D Compatible with E4D mills, same mandrel format as Planmill 3112x, 3114x, 4012x, 4014x. Lava Ultimate for Straumann milling centers 1334x. Lava Ultimate CAD/CAM Restorative for TS150, Compatible with TS150 (IOS Technologies), Glidewell Laboratories, 2614x. Lava Ultimate Restorative with Blue Mandrel, Pedestal Mandrel, compatible with Carestream CS 3000, Roland DWX-50, Roland DWX-4, 1314x, 2012x, 2110x. Lava Ultimate Restorative for Planmill, Compatible with Planmill mills, same mandrel format as E4D, 4212x, 4214x, 4312x, 4414x, 5072x, 5074x. Lava Ultimate Restorative for Ceramill, Compatible with Ceramill mills manufactured by Amann Girrbach, 4514x, 4714x, 5044x. Frame/Blank Version Used by Dental Laboratories: Lava Ultimate Implant Crown Restorative, Compatible with Lava Milling Equipment (Lava blocks bonded into frames. These are also known as blanks.), 69140-69149, 69150-69159, 69160-69163.