3M Healthcare Business recalls 3M Steri-Drape, Small Drape with Incise Film, REF 1021NS, general surgery drape
- Recall date
- October 6, 2022
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0256-2023
- FDA classification
- Class II
- Brand / firm
- 3M Healthcare Business
- Sold / distributed
- Worldwide distribution.
Why it was recalled
During a recent investigation, 3M confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Additionally, 3M has observed an increase in reported adhesive related skin injuries for these affected lots.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
3M Steri-Drape, Small Drape with Incise Film, REF 1021NS, general surgery drape