Medical device recalls Moderate risk

89 NORTH recalls LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 Laser Science fluorescence imaging

Recall date
January 31, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2820-2020
FDA classification
Class II
Brand / firm
89 NORTH
Sold / distributed
Worldwide

Why it was recalled

The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI; specifically, the potential for the LDI to enter a service mode intended for use in the factory environment only which overrides all interlocks. When this mode - "Interlock Override Mode" - is entered, the system does not cease laser emission when interlocks are opened, but rather allows emission to continue with open interlocks. The potential for this to occur is non-compliant with 21CFR 1040 .10{f )(2) and 21 CFR 1040 .10(f)(3).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 Laser Science fluorescence imaging

Get recall alerts

Free email alert whenever 89 NORTH has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: 89 NORTH