89 NORTH recalls LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 Laser Science fluorescence imaging
- Recall date
- January 31, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2820-2020
- FDA classification
- Class II
- Brand / firm
- 89 NORTH
- Sold / distributed
- Worldwide
Why it was recalled
The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI; specifically, the potential for the LDI to enter a service mode intended for use in the factory environment only which overrides all interlocks. When this mode - "Interlock Override Mode" - is entered, the system does not cease laser emission when interlocks are opened, but rather allows emission to continue with open interlocks. The potential for this to occur is non-compliant with 21CFR 1040 .10{f )(2) and 21 CFR 1040 .10(f)(3).
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 Laser Science fluorescence imaging
Get recall alerts
Free email alert whenever 89 NORTH has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: 89 NORTH