AB Sciex, LLC recalls Citrine Triple Quad Mass Spectrometer, Part Number 5063684
- Recall date
- August 30, 2023
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0040-2024
- FDA classification
- Class II
- Brand / firm
- AB Sciex, LLC
- Sold / distributed
- US Nationwide.
Why it was recalled
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Citrine Triple Quad Mass Spectrometer, Part Number 5063684
Get recall alerts
Free email alert whenever AB Sciex, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: AB Sciex, LLC