Medical device recalls Moderate risk

Abbott Gmbh & Co. KG recalls Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number A-30109694-01 Produ…

Recall date
December 20, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0789-2019
FDA classification
Class II
Brand / firm
Abbott Gmbh & Co. KG
Sold / distributed
Worldwide Distribution

Why it was recalled

Abbott has identified an issue with the Alinity i Gear Pump Assembly (part number A 30108552 01) resulting in foaming/bubbling out of the bottle reservoir for Concentrated Wash Buffer and an unexpected amount of dried residue of buffer. The foaming/bubbling is a result of an improper calibration of the pump s operating speed.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number A-30109694-01 Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.

Get recall alerts

Free email alert whenever Abbott Gmbh & Co. KG has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Abbott Gmbh & Co. KG