Abbott Ireland Diagnostics Division recalls ARCHITECT Free T3 Reagent Kit used on ARCHITECT i1000SR or i2000/i2000SR platforms Product Usage: It is a Chemiluminesc…
- Recall date
- October 12, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0798-2019
- FDA classification
- Class II
- Brand / firm
- Abbott Ireland Diagnostics Division
- Sold / distributed
- Worldwide Distribution - US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, PR. International distribution to AFGHANISTAN, ALBANIA, ALGERIA, ANG…
Why it was recalled
Samples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assays may show depressed results due to a reagent carryover when testing on board with specific assays on the ARCHITECT i1000SR and i2000/i2999SR platforms.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
ARCHITECT Free T3 Reagent Kit used on ARCHITECT i1000SR or i2000/i2000SR platforms Product Usage: It is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of free triiodothyronine (Free T3) in human serum and plasma.
Get recall alerts
Free email alert whenever Abbott Ireland Diagnostics Division has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Abbott Ireland Diagnostics Division