Medical device recalls Moderate risk

Abbott Laboratories, Inc recalls APS Accelerator modular systems, 208-230 V-, 50/60 Hz, 1800 VA, IVD, RX, Class 2 Laser Product,

Recall date
September 7, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2755-2019
FDA classification
Class II
Brand / firm
Abbott Laboratories, Inc
Sold / distributed
US: TX, VA, AZ, NV, MA, LA, MN, PA, SC, NJ, NY, CA, IL, IN, WA, FL, TN, PA, KS, ND, NC, CT OUS: Argentina, Australia, Belgium, Canada, China, Denmark, France, Greece, Hong Kong, Ireland, Italy, New Zealand, Norway, Oman, Pakistan, Philippines, Portugal, Russia, Saudi Arabia, Singapore, Spain, Taiwa…

Why it was recalled

The firm identified customer sites where modules which the safety pacemaker label is missing. Information regarding the label included in the Operations Manual has been updated to specify: 1) the risk of malfunction applies to anyone with a pacemaker who may work on these modules; 2) the minimum safety distance has increased to 200 mm (7.87") from transport mechanisms. Any person with a pacemaker must not get closer than this distance from modules axes.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

APS Accelerator modular systems, 208-230 V-, 50/60 Hz, 1800 VA, IVD, RX, Class 2 Laser Product,