Abbott Laboratories recalls ARCHITECT BNP Controls Ptoduct Usage: The ARCHITECT BNP Controls are for the estimation of test precision and the detec…
- Recall date
- May 24, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2079-2019
- FDA classification
- Class III
- Brand / firm
- Abbott Laboratories
- Sold / distributed
- Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed worldwide to ALGERIA, ANDORRA, AUSTRIA,…
Why it was recalled
Presence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing.
Class III means the product violates regulations but is not likely to cause harm.
What was recalled
ARCHITECT BNP Controls Ptoduct Usage: The ARCHITECT BNP Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT iSystem (reagents, calibrators and instrument), when used for the quantitative determination of human B type natriuretic peptide (BNP) in human EDTA plasma.
Get recall alerts
Free email alert whenever Abbott Laboratories has a new recall — straight from official government data. Unsubscribe anytime.