Medical device recalls Moderate risk

Abbott Point Of Care Inc. recalls Abbott i-STAT CG8+ Cartridges intended for use in the in vitro quantification of potassium in arterial, venous, or capi…

Recall date
September 13, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0615-2019
FDA classification
Class II
Brand / firm
Abbott Point Of Care Inc.
Sold / distributed
WI

Why it was recalled

Inaccurate potassium results may be generated. When testing with control fluid and/or patient samples with low or reference range potassium concentration, elevated potassium results may be generated. When testing with samples with a high concentration of potassium, decreased potassium results may be generated

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Abbott i-STAT CG8+ Cartridges intended for use in the in vitro quantification of potassium in arterial, venous, or capillary whole blood. List Number/UDI# : 03P88-25 ,(01)10054749000163(17)181028(10)L18059

Get recall alerts

Free email alert whenever Abbott Point Of Care Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Abbott Point Of Care Inc.