Abbott Point Of Care Inc. recalls i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 0005…
- Recall date
- September 12, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0926-2018
- FDA classification
- Class II
- Brand / firm
- Abbott Point Of Care Inc.
- Sold / distributed
- Distributed in 48 States: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, Australia, Italy, Sweden, and Canada.
Why it was recalled
Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)
Get recall alerts
Free email alert whenever Abbott Point Of Care Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Abbott Point Of Care Inc.