Abbott Vascular recalls Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only
- Recall date
- May 1, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1687-2019
- FDA classification
- Class II
- Brand / firm
- Abbott Vascular
- Sold / distributed
- U.S.: FL, UT, IN, OK, TX, CO, LA, AZ, MO, AL, GA, CA, AR, KS, IL, HI, IA, KY, MA, DC, NY, NV, NM, NJ, MT, MD, NC, CT, MI, MN, OH, ND, NH, OR, PA, WI, WA, VA, WV, TN, SC, RI, DE, ME, NE, SD, MS, ID, VT. Foreign (OUS): AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, Canada, CROATIA, CZECH REPUBLIC, DENMARK, F…
Why it was recalled
Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in unexpected movement of clip arms. The inability to close and remove the device has lead to surgery and additional intervention.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only
Get recall alerts
Free email alert whenever Abbott Vascular has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Abbott Vascular