Medical device recalls Moderate risk

Abbott Vascular recalls Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT, MitraClip G4 CDS NTW REPAIR-MR Clinical/CDS0704-NTW, M…

Recall date
September 8, 2022
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0117-2023
FDA classification
Class II
Brand / firm
Abbott Vascular
Sold / distributed
Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, A…

Why it was recalled

An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT, MitraClip G4 CDS NTW REPAIR-MR Clinical/CDS0704-NTW, MitraClip G4 CDS XT REPAIR-MR Clinical/CDS0704-XT, MitraClip G4 CDS XTW REPAIR-MR Clinical/CDS0704-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT, TriClip G4 NT Delivery System CLIN/TCDS0303-NT, TriClip G4 NTW Delivery System CLIN/TCDS0303-NTW, TriClip G4 XT Delivery System CLIN/TCDS0303-XT, TriClip G4 XTW Delivery System CLIN/TCDS0303-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT. For cardiac procedures.