Accriva Diagnostics, Inc. recalls AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.

Recall date
September 27, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0279-2025
FDA classification
Class II
Brand / firm
Accriva Diagnostics, Inc.
Sold / distributed
US: AZ

Why it was recalled

Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.