Medical device recalls Moderate risk

Accuray Incorporated recalls CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010:…

Recall date
April 10, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1697-2019
FDA classification
Class II
Brand / firm
Accuray Incorporated
Sold / distributed
US & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand

Why it was recalled

There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010: CyberKnife VSI Robotic Radiosurgery System

Get recall alerts

Free email alert whenever Accuray Incorporated has a new recall — straight from official government data. Unsubscribe anytime.