Medical device recalls Moderate risk

ACTIM OY recalls Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane

Recall date
November 23, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1592-2024
FDA classification
Class II
Brand / firm
ACTIM OY
Sold / distributed
US Nationwide distribution in the state of Connecticut.

Why it was recalled

Interference testing has shown that the presence of personal lubricants in the vaginal specimen can interfere with the PROM test result by providing a false positive result.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane

Get recall alerts

Free email alert whenever ACTIM OY has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: ACTIM OY