Medical device recalls

Advanced Bionics, LLC recalls SoundWave Professional Suite Software 3.2, Ref CI-6055-014, SPN 058-023-32085, UDI: (01)07630016853227, Logical SoundWa…

Recall date
April 4, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1441-2020
FDA classification
Class III
Brand / firm
Advanced Bionics, LLC
Sold / distributed
US Nationwide distribution in the sates of AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, OH, OK, PA, SC, SD, TN, TX, VA, VT, WI

Why it was recalled

The manufacturer received complaints that customers were attempting to install the fitting software using the supplied USB drive, an error message was received, and installation failed. The failed installation has not impact to the current version of the software, and users are able to continue programming devices with the current software version.

Class III means the product violates regulations but is not likely to cause harm.

What was recalled

SoundWave Professional Suite Software 3.2, Ref CI-6055-014, SPN 058-023-32085, UDI: (01)07630016853227, Logical SoundWave Professional Suite 3.2, CE 0123 - Product Usage: Soundwave 3.2 is used by healthcare professionals to program Advanced Bionics sound processors for use with Advanced Bionics cochlear implants, a device is intended to restore a level of auditory sensation to children with profound sensorineural deafness via electrical stimulation of the auditory nerve.

Get recall alerts

Free email alert whenever Advanced Bionics, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Advanced Bionics, LLC