Medical device recalls Moderate risk

Advanced Instruments, LLC recalls OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measur…

Recall date
March 11, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1696-2024
FDA classification
Class II
Brand / firm
Advanced Instruments, LLC
Sold / distributed
Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic, France, Germany, Iceland, Netherlands, South Korea, Spain, Switzerland, Taiwan.

Why it was recalled

System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and urine. Model/Catalog Number: OsmoPRO MAX Software Version: 1.5.0.0