Medical device recalls Moderate risk

Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocesso…

Recall date
May 11, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1714-2015
FDA classification
Class II
Brand / firm
Advanced Sterilization Products
Sold / distributed
Worldwide Distribution-US (nationwide) and the countries of Australia, Singapore, Canada, and Thailand.

Why it was recalled

Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.