Advanced Sterilization Products recalls EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and Re…
- Recall date
- May 11, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1715-2015
- FDA classification
- Class II
- Brand / firm
- Advanced Sterilization Products
- Sold / distributed
- Worldwide Distribution-US (nationwide) and the countries of Australia, Singapore, Canada, and Thailand.
Why it was recalled
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.