Aesculap Implant Systems LLC recalls GP321R; ELAN 4 2-RING DIAMOND BURR COARSE D2.3; GP322R; ELAN 4 2-RING DIAMOND BURR COARSE D3.0; GP323R; ELAN 4 2-RING D…
- Recall date
- March 28, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2088-2017
- FDA classification
- Class II
- Brand / firm
- Aesculap Implant Systems LLC
- Sold / distributed
- US Nationwide distribution - CA, IL, NY, PA, TN, TX
Why it was recalled
Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
GP321R; ELAN 4 2-RING DIAMOND BURR COARSE D2.3; GP322R; ELAN 4 2-RING DIAMOND BURR COARSE D3.0; GP323R; ELAN 4 2-RING DIAMOND BURR COARSE D4.0; GP324R; ELAN 4 2-RING DIAMOND BURR COARSE D5.0; GP325R; ELAN 4 2-RING DIAMOND BURR COARSE D6.0; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.