Medical device recalls Moderate risk

Aesculap, Inc. recalls S4C Occiput Torque Wrench F/Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with a defined tor…

Recall date
July 7, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2326-2015
FDA classification
Class II
Brand / firm
Aesculap, Inc.
Sold / distributed
Distributed US (nationwide) and to Canada.

Why it was recalled

A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after reprocessing.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

S4C Occiput Torque Wrench F/Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with a defined torque 2.8 Nm for tightening the clamping screw in the screw head/hook head/occipital plate/transverse connector and other connectors.