Alliance Partners LLC recalls Nakoma-SL Anterior Cervical Plating System's Variable Single Barrel Drill Guide Product Usage: The Nakoma-SL ACP System…
- Recall date
- June 9, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1989-2015
- FDA classification
- Class II
- Brand / firm
- Alliance Partners LLC
- Sold / distributed
- US Nationwide Distribution in the states LA, TX, and PR.
Why it was recalled
Inadequate interference between the screw and the plate when screw implanted into the plate at angulation greater than 7 degrees can result in the screw advancing through the plate during implantation.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Nakoma-SL Anterior Cervical Plating System's Variable Single Barrel Drill Guide Product Usage: The Nakoma-SL ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The Nakoma ACP System is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusion in patients with the following indications: Degenerative Disk Disease, Trauma, Tumors, Deformities or curvatures, Psuedoarthrosis, Failed previous fusions, Spondylolistesis, and Spinal Stenosis.
Get recall alerts
Free email alert whenever Alliance Partners LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Alliance Partners LLC