Ameditech Inc recalls Fastect¿ II Drug Screen Dipstick Test Single Drug Detection (AMT Item Number/ AMT Description): MF25/ "Fastect¿ II Drug…
- Recall date
- August 18, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-3034-2020
- FDA classification
- Class II
- Brand / firm
- Ameditech Inc
- Sold / distributed
- US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Fastect¿ II Drug Screen Dipstick Test Single Drug Detection (AMT Item Number/ AMT Description): MF25/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10", QF09/"Fastect¿ II Drug Screen Dipstick Test, 4 Drug Detection: THC50+COC300+OPI300+BZO300", MF20/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: MTD300", MF21/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: TCA1000, MF23/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF23/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF18/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BZO300, MF18/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BZO300", QF12/ "Fastect¿ II Drug Screen Dipstick Test 4 Drug Detection: THC50+COC300+OPI300+MET500", MF10/ "Fastect¿ II Drug Screen Dipstick Test THC50", MF25/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10", MF16 /"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: MET500", MF23 /"Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF14/ "Fastect¿ II Drug Screen Dipstick Test AMP1000", MF11/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: COC300", MF23 / Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100", MF25/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10", MF25/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: BUP10", MF23/ "Fastect¿ II Drug Screen Dipstick Test Single Drug Detection: OXY100"
Get recall alerts
Free email alert whenever Ameditech Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Ameditech Inc