Ameditech Inc recalls TDDA Drugs of Abuse Cup(AMT Item/ AMT Description): TDDA-6MBAU-CUP /"TDDA Drugs of Abuse Cup AMP300/BZO200/COC300/MET30…
- Recall date
- August 18, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-3044-2020
- FDA classification
- Class II
- Brand / firm
- Ameditech Inc
- Sold / distributed
- US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Why it was recalled
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
TDDA Drugs of Abuse Cup(AMT Item/ AMT Description): TDDA-6MBAU-CUP /"TDDA Drugs of Abuse Cup AMP300/BZO200/COC300/MET300/OPI300/ THC50 with Cr, Ni, pH, Bl, S.G."
Get recall alerts
Free email alert whenever Ameditech Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Ameditech Inc