Medical device recalls Moderate risk

American Medical Systems Innovation Center - Silicon Valley recalls 1X1 Cleaving Tiles; The Fiber Cleaver is a reusable and sterilizable tool used to score the outer jacket on Endostat an…

Recall date
May 28, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2171-2015
FDA classification
Class II
Brand / firm
American Medical Systems Innovation Center - Silicon Valley
Sold / distributed
Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Isra…

Why it was recalled

validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

1X1 Cleaving Tiles; The Fiber Cleaver is a reusable and sterilizable tool used to score the outer jacket on Endostat and SureFlex Fibers. The tool is a one piece pencil-like design with a sharpened blade at the end. The Fiber Cleaver is a SureFlex TM Reusable Laser Lithotripsy fiber and EndoStat fiber accessory. The Fiber Cleaver does not come in direct contact with the patient.