American Medical Systems Innovation Center - Silicon Valley recalls Fiber Cleaver; Fiber Cleaver is also used to score the outer jacket on EndoStat Fibers (Aura XP) The Fiber Cleaver is a…
- Recall date
- May 28, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2170-2015
- FDA classification
- Class II
- Brand / firm
- American Medical Systems Innovation Center - Silicon Valley
- Sold / distributed
- Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Isra…
Why it was recalled
validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Fiber Cleaver; Fiber Cleaver is also used to score the outer jacket on EndoStat Fibers (Aura XP) The Fiber Cleaver is a reusable and sterilizable tool used to score the outer jacket on Endostat and SureFlex Fibers. The tool is a one piece pencil-like design with a sharpened blade at the end. The Fiber Cleaver is a SureFlex TM Reusable Laser Lithotripsy fiber and EndoStat fiber accessory. The Fiber Cleaver does not come in direct contact with the patient.