Angiodynamics, Inc. recalls Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600, Model # (UPN) H7879006000, STERILE --- This is only for…
- Recall date
- June 5, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0526-2018
- FDA classification
- Class II
- Brand / firm
- Angiodynamics, Inc.
- Sold / distributed
- Worldwide distribution - US Nationwide
Why it was recalled
The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600, Model # (UPN) H7879006000, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.
Get recall alerts
Free email alert whenever Angiodynamics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Angiodynamics, Inc.