Medical device recalls Moderate risk

Angiodynamics, Inc. recalls Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601-US, Model # (UPN) H787900601US0, STERILE --- This is only for…

Recall date
June 5, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0530-2018
FDA classification
Class II
Brand / firm
Angiodynamics, Inc.
Sold / distributed
Worldwide distribution - US Nationwide

Why it was recalled

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601-US, Model # (UPN) H787900601US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.