Angiodynamics, Inc. recalls AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012
- Recall date
- June 27, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2444-2019
- FDA classification
- Class II
- Brand / firm
- Angiodynamics, Inc.
- Sold / distributed
- The products were distributed to the following US states: DC, IL, IN, MI, NJ, TN, TX, and WI.
Why it was recalled
The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012
Get recall alerts
Free email alert whenever Angiodynamics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.