Aomori Olympus Co., Ltd. recalls Olympus Distal Attachment (in Olympus Disposable EMR Kit)- Designed to be attached to the distal end of the endoscope.…
- Recall date
- July 27, 2023
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2522-2023
- FDA classification
- Class II
- Brand / firm
- Aomori Olympus Co., Ltd.
- Sold / distributed
- US Nationwide distribution.
Why it was recalled
The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Olympus Distal Attachment (in Olympus Disposable EMR Kit)- Designed to be attached to the distal end of the endoscope. Disposable EMR Kit: K-001, K-002, K-003, K-004, K-011, K-012