Medical device recalls High risk

Arkray Factory USA, Inc. recalls arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Ure…

Recall date
December 18, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0631-2016
FDA classification
Class I
Brand / firm
Arkray Factory USA, Inc.
Sold / distributed
Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.

Why it was recalled

One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.