Medical device recalls Moderate risk

AROA Biosurgery, LTD. recalls OviTex Reinforced BioScaffold 6x10cm, Part Number F10254-0610G

Recall date
April 6, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0416-2019
FDA classification
Class II
Brand / firm
AROA Biosurgery, LTD.
Sold / distributed
The products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.

Why it was recalled

Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

OviTex Reinforced BioScaffold 6x10cm, Part Number F10254-0610G

Get recall alerts

Free email alert whenever AROA Biosurgery, LTD. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: AROA Biosurgery, LTD.