Medical device recalls Moderate risk

Autochair recalls Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Recall date
March 7, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1420-2018
FDA classification
Class II
Brand / firm
Autochair
Sold / distributed
PA, and worldwide

Why it was recalled

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B