Bard Access Systems, Inc. recalls BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK000196, 917411…
- Recall date
- March 11, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1507-2025
- FDA classification
- Class I
- Brand / firm
- Bard Access Systems, Inc.
- Sold / distributed
- Worldwide - US Nationwide distribution including in the states of WI, NC, TX, NY, CA, MN, MT, RI, MA, OH, CT, SC, IL, CO, VA, NJ, GA, UT, FL, PA, IN, MD, KY, AZ, MI, WA, HI, MO, VT, AL, NM, TN, DE, LA, IA, NH, ID, DC, WV, KS, NE, SD, ND, WY, AK, OK, OR, AR and the countries of Andorra, Australia, A…
Why it was recalled
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK000196, 9174118, CK000955, 3174135, CK000617, CK000839, 3194355, 6174118, 3194135, 3194335, 2174108, 9194118, 3194118, 1174118, CK000285C, CK000335A, CK000796, 1194108D, CK000757, CK001042, 9174108, 1174108D1, CK000631B, 3194155, 3174108, CK000567, 1194108D4, 1174108D2, CK000677, CK000864, CK000937A, CK000930, CK001027, 1174108D, CK000867, CK000900, CK000315A, CK000589A, CK000854B, CK000897, CK000953A, CK000586B, CK000549, CK000844, 1194108D1, 1174108D4, 1194108D3, CK000934, CK000986, CK000411, CK001018, 1174108, CK000060, CK000513A, CK001021, CK001030, CK000501A, CK000619A, CK000903, 3194108, 1194108D2, 9194108D, CK000959, CK000613, CK000991, CK000983, CK000881, CK000721, CK000642, CK000289C, CK000894A, CK000914, CK000445, CK000687, CK000524A, CK000819, CK000266A, 1174108D3, 3194108D, CK000545, 1174108D5, 3174108D, CK000775A, 1194108D5, CK000736, CK000739, 9194108, 1194108, CK000660, CK000609A, 9174108D, CK001067, CK000077D, CK000882, CK000874, CK000994, CK000870A, CK000443C, CK000668B, CK000784
Get recall alerts
Free email alert whenever Bard Access Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Bard Access Systems, Inc.