Why it was recalled
In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to mitigate against Unintended Increased Intraperitoneal Volume (IIVP). This can cause serious problems in patients with unrelated a co-morbid condition of ascites.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. The HomeChoice/HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.