Medical device recalls Moderate risk

Baxter Healthcare Corp. recalls 1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2)…

Recall date
October 7, 2013
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2609-2016
FDA classification
Class II
Brand / firm
Baxter Healthcare Corp.
Sold / distributed
Worldwide distribution. US nationwide including Puerto Rico; Australia, Canada, New Zealand, and Japan

Why it was recalled

In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to mitigate against Unintended Increased Intraperitoneal Volume (IIVP). This can cause serious problems in patients with unrelated a co-morbid condition of ascites.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. The HomeChoice/HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.