Baxter Healthcare Corp. recalls V6 Rear Case Assembly, Product Code 35701, a replacement part for the SIGMA Spectrum Infusion Pump -- SIGMA LLC Medina,…
- Recall date
- September 28, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0672-2017
- FDA classification
- Class II
- Brand / firm
- Baxter Healthcare Corp.
- Sold / distributed
- Nationwide Distribution, Puerto Rico and Canada
Why it was recalled
A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
V6 Rear Case Assembly, Product Code 35701, a replacement part for the SIGMA Spectrum Infusion Pump -- SIGMA LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended. Product Code 35701 is a replacement assembly that is shipped to customers for self-repair of the Spectrum Infusion Pump. It is not a registered finished good.
Get recall alerts
Free email alert whenever Baxter Healthcare Corp. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Baxter Healthcare Corp.