Baxter Healthcare Corporation recalls Baxter EXACTAMIX Inlet, Vented, Syringe Inlet, REF H938176, for use in a pharmaceutical compounding device
- Recall date
- August 20, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-3085-2024
- FDA classification
- Class I
- Brand / firm
- Baxter Healthcare Corporation
- Sold / distributed
- Worldwide distribution.
Why it was recalled
Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
Baxter EXACTAMIX Inlet, Vented, Syringe Inlet, REF H938176, for use in a pharmaceutical compounding device
Get recall alerts
Free email alert whenever Baxter Healthcare Corporation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Baxter Healthcare Corporation