Medical device recalls High risk

Baxter Healthcare Corporation recalls Baxter EXACTAMIX Inlet, Vented, Syringe Inlet, REF H938176, for use in a pharmaceutical compounding device

Recall date
August 20, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-3085-2024
FDA classification
Class I
Brand / firm
Baxter Healthcare Corporation
Sold / distributed
Worldwide distribution.

Why it was recalled

Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Baxter EXACTAMIX Inlet, Vented, Syringe Inlet, REF H938176, for use in a pharmaceutical compounding device

Get recall alerts

Free email alert whenever Baxter Healthcare Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Baxter Healthcare Corporation