Medical device recalls Moderate risk

Baxter Healthcare Corporation recalls Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2

Recall date
January 26, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1693-2017
FDA classification
Class II
Brand / firm
Baxter Healthcare Corporation
Sold / distributed
Natiowide, Puerto Rico and Canada

Why it was recalled

SIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under conditions in which a message is being written to the Spectrum Infusion Pumps Event History Log (EHL), and the length of that message exactly matches the size of the available memory at the end of a memory sector. For this reason, the issue is significantly more likely to manifest when poor network maintenance results in high volumes of messages being written to the EHL.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2

Get recall alerts

Free email alert whenever Baxter Healthcare Corporation has a new recall — straight from official government data. Unsubscribe anytime.