Baxter Healthcare Corporation recalls SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009
- Recall date
- September 18, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0107-2019
- FDA classification
- Class II
- Brand / firm
- Baxter Healthcare Corporation
- Sold / distributed
- NJ
Why it was recalled
If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009
Get recall alerts
Free email alert whenever Baxter Healthcare Corporation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Baxter Healthcare Corporation