Baxter Healthcare Corporation recalls TherMax Blood Warmer Unit
- Recall date
- September 19, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0247-2020
- FDA classification
- Class II
- Brand / firm
- Baxter Healthcare Corporation
- Sold / distributed
- Nationwide distribution to Georgia and Maryland. International distribution to Belgium, Germany, Finland, France, Greece, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom, Australia, and New Zealand
Why it was recalled
TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
TherMax Blood Warmer Unit
Get recall alerts
Free email alert whenever Baxter Healthcare Corporation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Baxter Healthcare Corporation