Medical device recalls Moderate risk

Baxter Healthcare Corporation recalls TherMax Blood Warmer Unit

Recall date
September 19, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0247-2020
FDA classification
Class II
Brand / firm
Baxter Healthcare Corporation
Sold / distributed
Nationwide distribution to Georgia and Maryland. International distribution to Belgium, Germany, Finland, France, Greece, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom, Australia, and New Zealand

Why it was recalled

TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

TherMax Blood Warmer Unit

Get recall alerts

Free email alert whenever Baxter Healthcare Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Baxter Healthcare Corporation