Medical device recalls Moderate risk

BD LIFE SCIENCES recalls Software version 4.44A utilized on the BD MAX System

Recall date
July 29, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1864-2017
FDA classification
Class II
Brand / firm
BD LIFE SCIENCES
Sold / distributed
Worldwide Distribution - US Distribution to the states of : CA, CO, FL, IL, MN, MO, MT, NE, NJ, NY, OH, OR, PA, TX, TN, WA, WI., and to the countries of : Canada, Hong Kong, Taiwan, Turkey and Europe. Federal Government Agency sales/distribution centers or foreign countries VA Fort Harrison and VAM…

Why it was recalled

The BD MAX System software version 4.44A contains a software anomaly that has the potential to affect some Open System Reagent (OSR) customers utilizing User Defined Protocol (UDP) assays. Customers utilizing only IVD assays are not affected. This anomaly may cause the OSR customer system to incorrectly switch the columns of the truth table on the users display. If the user tries to fix their display data, the display will appear correct however, the database may become incorrect, potentially resulting in incorrect results reporting. BD is able to trace the issue to a limited number of instruments that have software version 4.44A installed. Note that the problem only affects customers that are either creating a new UDP with Result Logic or modifying the Result Logic of an existing UDP.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Software version 4.44A utilized on the BD MAX System

Get recall alerts

Free email alert whenever BD LIFE SCIENCES has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: BD LIFE SCIENCES