Medical device recalls Moderate risk

Beckman Coulter Inc. recalls Access BR Monitor, Part No. 387620

Recall date
April 13, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2074-2017
FDA classification
Class II
Brand / firm
Beckman Coulter Inc.
Sold / distributed
Affected geographies: Algeria, Australia, Austria, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Hong Kong,…

Why it was recalled

Following biases observed in the results of the French National Quality Controls, Beckman Coulter has been informed that a study completed by the French Competent Authority Agence nationale de s¿curit¿ du Medicament et des produits de sant¿ (ANSM) showed diagnosed cancer patients in non-remission of their metastatic disease, may have results below the upper reference limit (URL) of 31.3 U/mL as stated in the Access BR Monitor instructions for use (IFU).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Access BR Monitor, Part No. 387620

Get recall alerts

Free email alert whenever Beckman Coulter Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Beckman Coulter Inc.