Medical device recalls Moderate risk

Beckman Coulter Inc. recalls Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

Recall date
January 2, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1112-2025
FDA classification
Class II
Brand / firm
Beckman Coulter Inc.
Sold / distributed
Worldwide distribution - US Nationwide and the countries of Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Kazakhstan, Korea, Republic of, Malaysia, Mexico, Peru, Philip…

Why it was recalled

Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

Get recall alerts

Free email alert whenever Beckman Coulter Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Beckman Coulter Inc.