Medical device recalls Moderate risk

Beckman Coulter Inc recalls Direct Bilirubin Reagents (DBIL 2 x 200 ) and (DBIL 2 x 300), Ref:439715 (200 tests/cartridge) and 476856 (300 tests/ca…

Recall date
August 15, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1425-2020
FDA classification
Class II
Brand / firm
Beckman Coulter Inc
Sold / distributed
Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico,…

Why it was recalled

It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Direct Bilirubin Reagents (DBIL 2 x 200 ) and (DBIL 2 x 300), Ref:439715 (200 tests/cartridge) and 476856 (300 tests/cartridge), For In Vitro Diagnostic Use, Rx Only, UDI: 15099590575120(200 tests/cartridge) and 15099590625603(300 tests/cartridge)