Medical device recalls Moderate risk

Beckman Coulter Inc. recalls iChem VELOCITY Automated Urine Chemistry System, Catalog #700-7176-001 and iChem VELOCITY Automated Urine Chemistry Sys…

Recall date
March 29, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1894-2018
FDA classification
Class II
Brand / firm
Beckman Coulter Inc.
Sold / distributed
US Distribution to states of: AL, IA, GA, and FL.

Why it was recalled

The Strip Provider Modules (SPM) on some of the instruments had not been inspected during manufacture for correct seal placement and integrity.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

iChem VELOCITY Automated Urine Chemistry System, Catalog #700-7176-001 and iChem VELOCITY Automated Urine Chemistry System Computerless, Catalog #700-7177-001. The iChemVELOCITY automated urine chemistry system is an in-vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, as well as in a iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems.