Beckman Coulter Inc. recalls iQ200 Series Urine Microscopy Analyzer- All part numbers
- Recall date
- March 30, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1945-2020
- FDA classification
- Class II
- Brand / firm
- Beckman Coulter Inc.
- Sold / distributed
- US Distribution: Nationwide Foreign Distribution: Worldwide
Why it was recalled
A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
iQ200 Series Urine Microscopy Analyzer- All part numbers