Beckman Coulter Inc. recalls Kaluza C Flow Cytometry Software, RX Only in the U.S.A., Beckman Coulter, Inc. for the following software devices: (A)…
- Recall date
- September 17, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1982-2019
- FDA classification
- Class II
- Brand / firm
- Beckman Coulter Inc.
- Sold / distributed
- US: CA and NY OUS: United Kingdom, Switzerland and Germany
Why it was recalled
Compatibility issue between the software and Microsoft updates to Windows 7, 8, and 10. The compatibility issue causes the software to be unusable which could result in a delay of reporting results.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Kaluza C Flow Cytometry Software, RX Only in the U.S.A., Beckman Coulter, Inc. for the following software devices: (A) Kaluza C Perpetual License (B) Kaluza C Single Use 1 Year License (C) Kaluza C 5 User Network (D) Kaluza C 10 User Network (E) Kaluza C Perpetual Educational License (F) Kaluza C Single User 1 Year educational License (G) Kaluza C 10 User Network Educational License (H) Kaluza C 5 User Network Educational License (I) Kaluza C Flow Cytometry Software
Get recall alerts
Free email alert whenever Beckman Coulter Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Beckman Coulter Inc.