Medical device recalls Moderate risk

Beckman Coulter Inc. recalls Kaluza C Flow Cytometry Software, RX Only in the U.S.A., Beckman Coulter, Inc. for the following software devices: (A)…

Recall date
September 17, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1982-2019
FDA classification
Class II
Brand / firm
Beckman Coulter Inc.
Sold / distributed
US: CA and NY OUS: United Kingdom, Switzerland and Germany

Why it was recalled

Compatibility issue between the software and Microsoft updates to Windows 7, 8, and 10. The compatibility issue causes the software to be unusable which could result in a delay of reporting results.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Kaluza C Flow Cytometry Software, RX Only in the U.S.A., Beckman Coulter, Inc. for the following software devices: (A) Kaluza C Perpetual License (B) Kaluza C Single Use 1 Year License (C) Kaluza C 5 User Network (D) Kaluza C 10 User Network (E) Kaluza C Perpetual Educational License (F) Kaluza C Single User 1 Year educational License (G) Kaluza C 10 User Network Educational License (H) Kaluza C 5 User Network Educational License (I) Kaluza C Flow Cytometry Software