Medical device recalls Moderate risk

Beckman Coulter Inc. recalls Lactate, Catalog No. OSR6x93 Product Usage: Lactate OSR6x93 is intended as a System reagent for the quantitative determ…

Recall date
April 13, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1931-2016
FDA classification
Class II
Brand / firm
Beckman Coulter Inc.
Sold / distributed
Worldwide Distribution - US Nationwide including Puerto Rico and the countries of: Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denma…

Why it was recalled

Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the following assays: Cholesterol OSR6X16, Triglyceride OSR6X118, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Lactate, Catalog No. OSR6x93 Product Usage: Lactate OSR6x93 is intended as a System reagent for the quantitative determination of L-Lactate in human plasma and cerebrospinal fluid (CSF) on Beckman Coulter AU analyzers.